5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

794

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Pause 2

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.

Input range

0 - 999999 s

Default value

0 s

5.6.3.5.2.4

KFT Ipol - Control parameters

Tab: Method

KFT Ipol Properties... Control parameters

Command name

Name of the command.

Entry

25 characters

End point

EP at

Measured value for the endpoint.

Input range

–2,000.0 - 2,000.0 mV

Default value

250.0 mV

Selection

off

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-
played. The User setting must be selected in order to be able to edit these
parameters.

Selection

slow | optimal | fast | User

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted EPs.
optimal
Parameter set for all standard titrations; optimized for the most fre-
quent applications.
fast
For less critical, fast titrations.