5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

617

Additional external measured values

Additional external measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can
be taken over and saved from existing measuring commands (i.e. MEAS)
which can also be displayed in curves under the designation Extern 1…3
and used in formulas as variables 'Command name.EX1…3'-

[New]

Open the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 934)
.

[Properties]

Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 934)
.

[Delete]

Delete the external measured value selected in the table.

5.6.3.3.3.9

MET U - Measured value window

Dialog window: Method

MET U Properties... Potentiometric evalua-

tion

Evaluation with measured value window [New]/[Properties] Mea-

sured value window#

Measured value windows are regions on the measured value axis for
which different parameters for potentiometric evaluation can be defined.
Only those endpoints which lie within these windows and additionally
meet the parameters defined for each window will be recognized. Only
one endpoint will be recognized per window.

Lower limit

Lower limit for the measured value window.

Titrando, 855
Input range

-2000.0 - 2000.0 mV

Default value

-2000.0 mV

Titrino
Input range

-2000 - 2000 mV

Default value

-2000 mV

Upper limit

Upper limit for the measured value window.