5 Method
tiamo 2.4 (for Process Analysis)
■■■■■■■■
749
Pause 2
Pause 2
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.
Input range
0 - 999999 s
Default value
0 s
5.6.3.4.5.4
SET Ipol - Control parameters
Tab: Method
▶ SET Ipol ▶ Properties... ▶ Control parameters
Command name
Name of the command.
Entry
25 characters
The control parameters can be set for each endpoint separately.
Endpoint 1
EP1 with
Measured value for the first endpoint.
Titrando, 855
Input range
-2000.0 - 2000.0 mV
Selection
off
Default value
off
Titrino
Input range
-2000 - 2000
Selection
off
Default value
off
Titration rate
Titration rate
The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-
played. The User setting must be selected in order to be able to edit these
parameters.
Selection
slow | optimal | fast | User
Default value
optimal