2 elt loop - properties – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1139

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1123

Identi-
fica-
tion

Description

.ETR

Result of the electrode test:

0 = No data available

1 = Short circuit

2 = Excessive start drift

3 = Wrong buffer

4 = Problems with the diaphragm

5 = Reference system defective

6 = No buffer values for the temperature or excessive devia-
tion of the temperature between buffers

7 = Partial short circuit

8 = Glass defective

9 = Poor electrode

10 = Usable electrode

11 = Good electrode

12 = Excellent electrode

.FIN

Command status; 1 = Command has ended at least
once
, 0, invalid (variable not available) = Command has
never ended

.LCO

Loop counter = current number of completed runs, both of
Repeat loops and of While loops

.LST

Start time of the loop command (Date/Time)

.SLO

Electrode slope of the sensor used for the command (in mV
for ISE sensors)

.VAR

Variance of the calculated result of the calibration in the
selected unit

5.6.5.8.2

ELT LOOP - Properties

Dialog window: Method

ELT LOOP Properties...

Command name

Name of the command.

Entry

25 characters