5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

844

■■■■■■■■

tiamo 2.4 (for Process Analysis)

5.6.3.6.10

KFC - Fixed endpoint evaluation

Dialog window: Method

KFC Properties... Additional evaluations Fixed

endpoint evaluation

[New]/[Property] Fixed endpoint evaluation #

Measured quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Water

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value
Input range

-2000.0 - 2000.0 mV

Time
Input range

0.0 - 999999.9 s

Water
Input range

0.0 - 999999.9 µg

5.6.3.7

BRC

5.6.3.7.1

BRC - Overview

Dialog window: Method

BRC Properties... BRC - 'Command name'

Command for coulometric titration for determining the bromine con-
sumption. Out of the amount of bromine consumed the bromine index
(amount of bromine consumed (in mg) per 100 g sample) and the bro-
mine number
(amount of bromine (in g) per 100 g sample) can be calcu-
lated.

Instruments
This command can be executed with the following devices.

Titrando: 851, 852

Appearance
The command has the following appearance:

without conditioning

with conditioning