5 analog outputs, 1 analog outputs - overview – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1720

Advertising
background image

8.4 Manual control - Functions

■■■■■■■■■■■■■■■■■■■■■■

1704

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Graphical display

8.4.6.5

Analog outputs

8.4.6.5.1

Analog outputs - Overview

Program part: Manual control

If in the device tree the branch Analog outputs is selected, then all out-
puts can be commonly controlled on the tabs Set value and Set default
in the subwindow functions/parameters.

In the subwindow for the graphical display the adjacent value for all con-
nected analog outputs is shown in their physical order.

Setting a value

Value

Enter the value in % of the mapped range.

Input range

0.0 - 100.0 % Mapping

Default value

0.0 % Mapping

[Set]

Sets all analog outputs to the mapping value indicated.

Setting a default

[Set]

Resets all analog outputs to the default value in the configuration.

Graphical display
Example: