5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

928

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Additional external measured values

Additional external measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can
be taken over and saved from existing measuring commands (i.e. MEAS)
which can also be displayed in curves under the designation Extern 1…3
and used in formulas as variables 'Command name.EX1…3'-

[New]

Open the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 934)
.

[Properties]

Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 934)
.

[Delete]

Delete the external measured value selected in the table.

5.6.3.9.9

TET - Endpoint

Dialog window: Method

TET Properties... Thermometric evalua-

tion

[New]/[Properties] Endpoint

Reaction type

Selection of the reaction type according to which the curve is to be evalu-
ated.

Selection

exothermic | endothermic

Default value

exothermic

EP criterion

Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.

for reaction type = exothermic
Input range

-999999 - 0

Default value

-5

for reaction type = endothermic
Input range

0 - 999999

Default value

5