5 meas t/flow - additional measured values, Additional measured values – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1024

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1008

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated time for the minimum mea-
sured value will be interpolated from the measuring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Input range

0.1 - 20.0 °C/s

Default value

1.0 °C/s

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated time for the maximum mea-
sured value will be interpolated from the measuring point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Input range

0.1 - 20.0 °C/s

Default value

1.0 °C/s

5.6.4.8.5

MEAS T/Flow - Additional measured values

Tab: Method

MEAS T/Flow Properties... Additional measured values

Command name

Name of the command.

Entry

25 characters

A maximum of 6 additional measured values can be defined on this tab
which can then be saved together with the measured values present in
the default settings in additional measured value columns Calculated #
or External # in the measuring point list .

Additional calculated measured values

Additional calculated measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can
be calculated from existing measured values and saved, which can also be

Advertising