4 meas ipol - evaluations – Metrohm tiamo 2.4 (process analysis) User Manual
Page 989
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4 (for Process Analysis)
■■■■■■■■
973
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Titrando, 855, 867
Input range
0.1 - 999999.0 s (Increment: 0.1)
Default value
2.0 s
Titrino
Input range
0.08 - 16200 s (Increment: 0.08)
Default value
2.0 s
Stop measured value
Stops when the preset measured value has been reached since the start of
the measurement.
Titrando, 855, 867
Input range
-2000.0 - 2000.0 mV
Selection
off
Default value
off
off
No stop.
Temperature
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.
Titrando, 855, 867
Input range
-20.0 - 150 °C
Default value
25.0 °C
Titrino
Input range
-170.0 - 500 °C
Default value
25.0 °C
5.6.4.5.4
MEAS Ipol - Evaluations
Tab: Method
▶ MEAS Ipol ▶ Properties... ▶ Evaluations
Command name
Name of the command.
Entry
25 characters