4 meas ipol - evaluations – Metrohm tiamo 2.4 (process analysis) User Manual

Page 989

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

973

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Titrando, 855, 867
Input range

0.1 - 999999.0 s (Increment: 0.1)

Default value

2.0 s

Titrino
Input range

0.08 - 16200 s (Increment: 0.08)

Default value

2.0 s

Stop measured value

Stops when the preset measured value has been reached since the start of
the measurement.

Titrando, 855, 867
Input range

-2000.0 - 2000.0 mV

Selection

off

Default value

off

off
No stop.

Temperature

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.

Titrando, 855, 867
Input range

-20.0 - 150 °C

Default value

25.0 °C

Titrino
Input range

-170.0 - 500 °C

Default value

25.0 °C

5.6.4.5.4

MEAS Ipol - Evaluations

Tab: Method

MEAS Ipol Properties... Evaluations

Command name

Name of the command.

Entry

25 characters