6 Configuration

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1429

6.7.3.3

Properties - Calibration data

Tab: Configuration

Sensors [Edit] Properties... Calibration data

NOTE

This tab only appears for sensors of the type pH electrode, ISE elec-
trode
, Conductivity sensor and Thermoprobe.

Specific calibration data for pH electrodes

Slope

Slope of the pH electrode. This value can be determined automatically
with a calibration from the linearized calibration curve or entered man-
ually. The slope of a pH electrode is available for calculations as the varia-
ble 'Command name.SLO.

Input range

–999.9 - 999.9 %

Default value

100.0 %

pH(0)

Apart from the slope, pH(0) is the second characteristic of the calibration
function U = f(pH). pH(0) is the x-axis intercept of the calibration curve,
i.e., it corresponds to the pH value at 0 mV. The electrode zero point
pH(0) of a pH electrode is available for calculations as the variable 'Com-
mand name.ENP'
.

Input range

–20.000 - 20.000

Default value

7.000

Specific calibration data for ISE electrodes

Slope

Slope of the ISE electrode. This value can be determined automatically
with a calibration from the linearized calibration curve or entered man-
ually. The slope of an ISE electrode is available for calculations as the varia-
ble 'Command name.SLO.

Input range

–999.9 - 999.9 mV

Default value

59.2/valence mV

E(0)

E(0) is the second characteristic of the calibration function U = f(log c).
E(0) is the y-axis intercept of the calibration curve, i.e., it corresponds to
the potential at log c = 0. The electrode zero point E(0) of an ISE elec-
trode is available for calculations as the variable 'Command name.ENP'.