5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1065

Input range

0 - 999,999 s

Default value

600 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Input range

0.25 - 999,999.0 s

Default value

2.0 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Input range

0 - 4,000 mAU

Selection

off

Default value

off

Temperature (089 Photometer)

Temperature

Measuring temperature, which can be entered manually. If the Tempera-
ture measurement
is set to continuous on the General/Hardware
tab under Sensor, then the temperature will be measured continuously.

Input range

20.0 - 60 °C

Default value

25.0 °C

5.6.4.15.4

MEAS Opt Conc - Additional measured values

Tab: Method

MEAS Opt Conc Properties... Additional measured values

Command name

Name of the command.

Entry

25 characters

A maximum of 6 additional measured values can be defined on this tab.
These values can then be saved together in additional measured value col-
umns with the measured values present in the default settings.

Additional calculated measured values

Additional calculated measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can
be calculated from existing measured values and saved, which can also be
presented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.