5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

783

5.6.3.4.6.10

SET Upol - Fixed endpoint evaluation

Dialog window: Method

SET Upol Properties... Additional evalua-

tions

Fixed endpoint evaluation [New/[Properties] Fixed endpoint evalua-

tion #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value
Input range

-200.0 - 200.0 µA

Time
Input range

0.0 - 999999.9 s

Volume
Input range

0.00000 - 9999.99 mL

5.6.3.5

KFT

5.6.3.5.1

KFT - Overview

Menu item: Method

Insert New command...

Command for volumetric Karl Fischer titrations.

Principle
Volumetric titrations for determining the water content according to Karl
Fischer with automatic conditioning are carried out with this command. A
typical feature of the Karl Fischer titration is the controlled reagent dosing,
which depends on the control difference (current measured value – end-
point), with the aim of reaching the measured value given as being the
predefined target endpoint as quickly and accurately as possible. Overti-
trating is avoided to a great extent. The titration termination at the end-
point takes place either drift-controlled or after a waiting period. The vol-
ume added up to the endpoint gives the calculable reagent consumption.

Reagent dosing takes place in three phases during the titration:

Initial dosing
The dosing rate increases continuously during this phase. It starts with
the Min. rate and increases to the Max. rate.