Using the 2-point shift method – Rockwell Automation 900-TC32 Digital Temperature Controllers, Series B User Manual

Page 141

Advertising
background image

Publication 900-UM007D-EN-E - January 2011

Parameter Adjustments & Application Considerations

4-5

Figure 4.12 — One-Point Temperature Input Shift

3. After you have configured the 1-point input shift value, check the

controller readout (A) and control target temperature (B). If they are
almost the same, this completes temperature input shift procedure.

Using the 2-Point Shift Method

Use the 2-point input shift method if you want to increase the accuracy of the
readout values across the range of your temperature sensor.

1. When using the 2-point shift method, shift the controller readout at two

points: — the first, near room temperature and — the second, near the
value at which the temperature of the control target is to be controlled.

Therefore, for this procedure, bring the control target temperature to
near room temperature and to near the set point, and check control
target temperature (B) and controller readout (A). Refer to Figure 4.10

2. Using the equations below, calculate the upper- and lower-limit

temperature input shift values from the readout and temperature to be
shifted that you obtained in step 1.

– Equation 1: Use the following equation to calculate the Lower-Limit

temperature input shift value.

– Equation 2: Use the following equation to calculate the Upper-Limit

temperature input shift value.

– Figure – shows the effect of shift by 2-point temperature input shift.

Note: For meanings of variables, refer to Figure –.

Controller readout (A)

Temperature readout
after shift (e.g. 120

°

C)

Temperature readout
before shift (e.g. 110

°

C)

After shift

Input shift value (e.g. 10

°

C)

Before shift

Near set point
(e.g. 120

°

C)

Temperature readout of
control target (B)

0

insl

=

YL

Y1

Y2

Y1

×

{(X2

Y2)

(X1

Y1)} + (X1

Y1)

insh

=

YH

Y1

Y2

Y1

×

{(X2

Y2)

(X1

Y1)} + (X1

Y1)

Advertising