6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Save

This menu is the same as the Select Trace menu under Trace. The trace selected on that menu
appears selected here, and selecting a trace here causes the same trace to be selected on the Select
Trace menu. (That is, there is only one "selected trace".)

The next step in the Save process is to select the Save As key in the Export Data menu.

Key Path

Save, Data, Trace

Notes

auto return

Couplings

When you select which trace to save, it makes that trace the current trace, so it displays on top
of all of the other traces.

Preset

Not part of Preset, but is reset to TRACE1 by Restore Mode Defaults; survives shutdown

State Saved

The current trace number is saved in instrument state.

Initial S/W Revision

Prior to A.02.00

Limit

Pressing this key selects Limit Lines as the data type to be exported. Pressing the key a second
time brings up the Limit Menu that allows you to select which Limit Line to save.

See

"Limits File Contents" on page 1055

.

See

".csv file format" on page 1055

See

".lim file format" on page 1056

Key Path

Save, Data

Remote Command

:MMEMory:STORe:LIMit LLINE1|LLINE2,<filename>

Example

:MMEM:STOR:LIM LLINE2, "myLimitLine2.csv" Saves the 2nd Limit Line to the file
myLimitLine2.csv in the current path.  The default path is My Documents\SA\data\limits

Notes

If the save is initiated via SCPI, and the file already exists, the file will be overwritten.

Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during an
instrument software upgrade.

Both single and double quotes are supported for any filename parameter over SCPI.

Dependencies

This key will only appear if you have the proper option installed in your instrument.

Preset

1; not part of Preset, but is reset by Restore Mode Defaults and survives power cycles

State Saved

The selected Limit number is saved in instrument state.

Readback

selected Limit Line

Status Bits/OPC
dependencies

Sequential - waits for previous measurement to complete

Initial S/W Revision

A.02.00