Agilent Technologies Signal Analyzer N9030a User Manual

Page 635

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Input/Output

If we interpolate on a linear scale, we assume that the two points are connected by a straight line
on the linear scale, as below:

The correction to be used for each bucket is taken from the interpolated correction curve at the
center of the bucket.

Description

Sets an ASCII description field which will be stored in an exported file.  Can be displayed in the
active function area by selecting as the active function, if desired to appear in a screen capture.

Key Path

Input/Output, Corrections, Properties

Remote Command

[:SENSe]:CORRection:CSET[1]|2|...6:DESCription "text"

[:SENSe]:CORRection:CSET[1]|2|...6:DESCription?

Example

:CORR:CSET1:DESC "11941A Antenna correction"

Notes

45 chars max; may not fit on display if max chars used

Preset

Unaffected by a Preset. Set to empty by Restore Input/Output Defaults

State Saved

Saved in instrument state.

Initial S/W Revision

A.02.00

Comment

Sets an ASCII comment field which will be stored in an exported file.  Can be displayed in the
active function area by selecting as the active function, if desired to appear in a screen capture.

Key Path

Input/Output, Corrections, Properties

Remote Command

[:SENSe]:CORRection:CSET[1]|2|...6:COMMent "text"

[:SENSe]:CORRection:CSET[1]|2|...6:COMMent?

Remote Language Compatibility Measurement Application Reference

635