6  RLC Swept SA Measurement Front-Panel & SCPI Reference
AMPTD Y Scale

trace 1 is on the fifth graticule line from the top, it represents 5 dBm and will read out remotely
as 5.

Example 2, set the following:

Scale Type (Lin)

Y Axis Unit, Volts

Ref Level, 100 mV (10 mV/div)

This sets the top line to 100 mV and the bottom line to 0 V, so each vertical division represents 10
mV. Thus, if a point on trace 1 is on the fifth graticule line from the top, it represents 50 mV and
will read out remotely as 50.

Dependencies

If an amplitude correction with an Antenna Unit other than None is applied and enabled, then
that antenna unit is forced and the key with that unit is the only Y Axis Unit available. All other Y
Axis Unit keys are grayed out. 

If an amplitude correction with an Antenna Unit other than None is applied and enabled, and you
then turn off that correction or set Apply Corrections to No, the Y Axis Unit that existed before
the Antenna Unit was applied is restored.

Couplings

The analyzer retains the entered Y Axis Unit separately for both Log and Lin amplitude scale
types

Preset

dBm for log scale, V for linear. The true ‘preset’ value is dBm, since at preset the Y Scale type is
set to logarithmic.

State Saved

Saved in instrument state

Readback line

1-of-N selection

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.02.00, A.04.00, A.11.00

dBm

Sets the amplitude unit for the selected amplitude scale (log/lin) to dBm.

Key Path

AMPTD Y Scale, Y Axis Unit

Example

UNIT:POW DBM

Dependencies

Grayed out if an Amplitude Correction with an Antenna Unit is ON. 

Readback

dBm

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.02.00

534

Remote Language Compatibility Measurement Application Reference