Agilent Technologies Signal Analyzer N9030a User Manual

Page 144

Advertising
background image

3  Programming the Analyzer
STATus Subsystem

"Questionable Negative Transition " on page 145

"Questionable Positive Transition" on page 146

Questionable Condition

This query returns the decimal value of the sum of the bits in the Questionable Condition register.

The data in this register is continuously updated and reflects the current conditions.

Mode

All

Remote Command

:STATus:QUEStionable:CONDition?

Example

STAT:QUES:COND?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Enable

This command determines which bits in the Questionable Event register will set the Questionable
Status Summary bit (bit3) in the Status Byte Register. The variable <integer> is the sum of the
decimal values of the bits you want to enable.

The preset condition is all bits in this enable register set to 0. To have any Questionable Events reported to
the Status Byte Register, one or more bits need to be set to 1. The Status Byte Event Register should be
queried after each measurement to check the Questionable Status Summary (bit 3). If it is equal to 1, a
condition during the test may have made the test results invalid. If it is equal to 0, this indicates that no
hardware problem or measurement problem was detected by the analyzer.

Mode

All

Remote Command

:STATus:QUEStionable:ENABle 16

:STATus:QUEStionable:ENABle?

Example

STAT:OPER:PTR 1 Align Now operation beginning will be reported to the Status Byte Register.

Preset

0

Min

0

Max

32767

Status Bits/OPC

Sequential command