6  RLC Swept SA Measurement Front-Panel & SCPI Reference
System

24 hours

With Auto Align Alert set to 24 Hours the instrument will signal an alert after a time span of 24
hours since the last successful full alignment (for example, Align Now, All or completion of a full
Auto Align). You may choose this selection in an environment where the temperature is stable on
a daily basis at a small risk of accuracy errors in excess of the warranted specifications. The alert
is the Error Condition message “Align Now, All required”.

For front-panel operation , confirmation is required  to transition into this setting of Alert. The
confirmation dialog is:

No confirmation is required when Alert is configured through a remote command.

Key Path

System, Alignments, Auto Align, Alert

Mode

All

Example

:CAL:AUTO:ALER DAY

Readback Text

24 hours

Status Bits/OPC
dependencies

Bit 14 is set in the Status Questionable Calibration register.

Initial S/W Revision

Prior to A.02.00

7 days

With Auto Align Alert is set to 7 days the instrument will signal an alert after a time span of 168
hours since the last successful full alignment (for example, Align Now, All or completion of a full
Auto Align). You may choose this selection in an environment where the temperature is stable on
a weekly basis, at a modest risk of accuracy degradations in excess of warranted performance.
The alert is the Error Condition message “Align Now, All required”.

For front panel operation, confirmation is required for the customer to transition into this setting
of Alert. The confirmation dialog is:

1266

Remote Language Compatibility Measurement Application Reference