Agilent Technologies Signal Analyzer N9030a User Manual

Page 652

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output

Example

TRIG1:OUTP GATE

Readback

Gate

Initial S/W Revision

Prior to A.02.00

Odd/Even Trace Point

Selects either the odd or even trace points as the signal to be output to the Trig 1 Out or Trig 2
Out connector when performing swept spectrum analysis. When the polarity is positive, this
output goes high during the time the analyzer is sweeping past the first point (Point 0) and every
other following trace point. The opposite is true if the polarity is negative.

Key Path

Input/Output, Output Config, Trig 1/2 Output

Example

TRIG1:OUTP OEV

Readback

Odd/Even

Initial S/W Revision

Prior to A.02.00

Trig Out (1 and 2)

Select the type of output signal that will be output from the rear-panel Trig 1 Out or Trig 2 Out
connectors.

Key Path

Input/Output, Output Config

Remote Command

:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut HSWP | MEASuring | MAIN |
GATE | GTRigger | OEVen | SPOint | SSWeep | SSETtled | S1Marker |
S2Marker | S3Marker | S4Marker | OFF

:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut?

Example

TRIG:OUTP HSWP

TRIG2:OUTP GATE

Dependencies

The second Trigger output (Trig 2 Out) does not appear in all models; in models that do not
support it, the Trig 2 Out key is blanked, and sending the SCPI command for this output
generates an error, “Hardware missing; Not available for this model number”   In models that do
not support the Trigger 2 output, this error is returned if trying to set Trig 2 Out and a query of
Trig 2 Out returns OFF.

Preset

Trigger 1: Sweeping (HSWP)

Trigger 2: Gate

This is unaffected by a Preset but is preset to the above values on a "Restore Input/Output
Defaults" or "Restore System Defaults->All"

652

Remote Language Compatibility Measurement Application Reference