6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Mode Preset

Mode Preset

Returns the active mode to a known state.

Mode Preset does the following for the currently active mode:

l

Aborts the currently running measurement.

l

Brings up the default menu for the mode, with no active function.

l

Sets measurement Global settings to their preset values for the active mode only.

l

Activates the default measurement.

l

Brings up the default menu for the mode.

l

Clears the input and output buffers.

l

Sets Status Byte to 0.

Mode Preset does not:

l

Cause a mode switch

l

Affect mode persistent settings

l

Affect system settings

See

"How-To Preset" on page 884

for more information.

Key Path

Front-panel key

Remote Command

:SYSTem:PRESet

Example

:SYST:PRES

Notes

*RST

is preferred over :SYST:PRES for remote operation. *RST does a Mode Preset, as done by

the :SYST:PRES command, and it sets the measurement mode to Single measurement rather
than Continuous for optimal remote control throughput.

Clears all pending OPC bits. The Status Byte is set to 0.

Couplings

A Mode Preset  aborts the currently running measurement, activates the default measurement,
and.  gets the mode to a consistent state with all of the default couplings set.

Backwards Compatibility
Notes

In the X-Series, the legacy “Factory Preset” has been replaced with Mode Preset, which only
presets the currently active mode, not the entire instrument.  In the X-Series, the way in to
preset the entire instrument is by using System, Restore System Defaults All, which behaves
essentially the same way as restore System Defaults does on ESA and PSA.

There is also no “Preset Type” as there is on the PSA.   There is a green Mode Preset front-panel
key that does a Mode Preset and a white-with-green-letters User Preset front-panel key that

Remote Language Compatibility Measurement Application Reference