Scale / div, Scale type – Agilent Technologies Signal Analyzer N9030a User Manual

Page 529

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

AMPTD Y Scale

Scale / Div

Sets the units per vertical graticule division on the display. This function is only available when
Scale Type (Log) is selected and the vertical scale is power. When Scale Type (Lin) is selected,
Scale/Div is grayed out.

Key Path

AMPTD Y Scale

Remote Command

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:PDIVision <rel_ampl>

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:PDIVision?

Example

DISP:WIND:TRAC:Y:PDIV 5 DB

Dependencies

Scale/Div is grayed out in linear Y scale. Sending the equivalent SCPI command does change the
Scale/Div, though it has no affect while in Lin.

Preset

10.00 dB / Div

State Saved

Saved in instrument state

Min

0.10 dB

Max

20 dB

Initial S/W Revision

Prior to A.02.00

Scale Type

Chooses a linear or logarithmic vertical scale for the display and for remote data readout.

When Scale Type (Log) is selected, the vertical graticule divisions are scaled in logarithmic units.
The top line of the graticule is the Reference Level and uses the scaling per division Scale/Div to
assign values to the other locations on the graticule.

When Scale Type (Lin) is selected, the vertical graticule divisions are linearly scaled with the
reference level value at the top of the display and zero volts at the bottom. Each vertical division
of the graticule represents one-tenth of the Reference Level.

The Y Axis Unit used for each type of display is set by pressing Y Axis Unit. The analyzer remembers
separate Y Axis Unit settings for both Log and Lin.

Key Path

AMPTD Y Scale

Remote Command

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:SPACing LINear | LOGarithmic

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:SPACing?

Example

DISP:WIND:TRAC:Y:SPAC LOG

DISP:WIND:TRAC:Y:SPAC?

Remote Language Compatibility Measurement Application Reference

529