6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.

Key Path

Trigger, External 1

Remote Command

:TRIGger[:SEQuence]:EXTernal1:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:EXTernal1:SLOPe?

Example

TRIG:EXT1:SLOP NEG

Couplings

This same slope is used in the Ext1 selection for the trigger source in the Trigger menu and for
the period timer sync source (in the Trigger menu and in the Gate Source menu).

Preset

POSitive

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:EXTernal:SLOPe

For backward compatibility, the parameter EXTernal is mapped to EXTernal1

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:FRAMe:EXTernal1:SLOPe

Backwards Compatibility
Notes

The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.

Initial S/W Revision

Prior to A.02.00

External 2

Pressing this key, when it is not selected, selects the external input port that you will use for the
periodic frame trigger synchronization.

Pressing this key, when it is already selected, accesses the external 2 sync source setup menu.

Key Path

Trigger, Periodic Timer, Sync Source

Example

TRIG:FRAM:SYNC EXT2

Dependencies

In some models, there is no second External input.  In these models, the External 2 key is
blanked and the EXTernal2 parameter will generate a “Hardware missing; Not available for this
model number” message.

Couplings

Same as External 2 trigger source.

Readback

External 2

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.03.00

1424