5  Legacy Command Descriptions

Command Syntax

Syntax
Component

Definition/Range

Analyzer
command

Any command in this chapter, with required parameters and terminators.

Character

SP a b c d e f g h i j k l m n o p q r s t u v w x y z databyte.

Character
& EOI

8-bit byte containing only character data and followed by end-or-identify (EOI) condition, where the EOI control
line on GPIB is asserted to indicate the end of the transmission. END signifies the EOI condition.

Character
string

A list of characters.

Data byte

8-bit byte containing numeric or character data.

Data byte &
EOI

8-bit byte containing numeric or character data followed by end-or-identify (EOI) condition, where the EOI
control line on GPIB is asserted to indicate the end of the transmission. END signifies the EOI condition.

Delimiter

| \ @ ˆ $ % ; ! Matching characters that mark the beginning and end of a character string, or a list of
commands. Choose delimiting characters that are not used within the string they delimit.

Digit

0 1 2 3 4 5 6 7 8 9

lsb length

Represents the least significant byte of a two-byte word that describes the number of bytes returned or
transmitted. See msb length.

msb length

Represents the most significant byte of a two-byte word that describes the number of bytes returned or
transmitted. See lsb length.

Number

Expressed as integer, decimal, or in exponential (E) form.

Integer Number Range: –32,768 through +32,767

General formatting restrictions:

Real Number Range: ± 1.797693134862315 x 10

308

, including 0.

Up to 15 significant figures allowed.

Numbers may be as small as ± 2.225073858507202 x 10

–308

Output
termination

Line feed (LF) and end-or-identify (EOI) condition. ASCII code 10 (line feed) is sent via GPIB and the end-or-
identify control line on GPIB sets to indicate the end of the transmission.

Units

Represent standard scientific units:

Frequency Units: GZ, GHZ, MZ, MHZ, KZ, KHZ, HZ

Amplitude Units: DB, DBMV, DM, DBM, DBUV, V, MV, UV, W, MW, UW

Time Units: SC, S, MS, US

Syntax Elements