3  Programming the Analyzer
STATus Subsystem

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Calibration Extended Needed Negative Transition

This command determines which bits in the Questionable Calibration Extended Needed Condition
register will set the corresponding bit in the Questionable Calibration Extended Needed Event
register when the condition register bit has a negative transition (1 to 0). The variable <integer>
is the sum of the decimal values of the bits that you want to enable.

Mode

All

Remote
Command

:STATus:QUEStionable:CALibration:EXTended:NEEDed:NTRansition <integer>

:STATus:QUEStionable:CALibration:EXTended:NEEDed:NTRansition?

Example

STAT:QUES:CAL:EXT:NEED:NTR 2

Align EMI conducted is not required.

Preset

0

Min

0

Max

32767

Status Bits/OPC
dependencies

Sequential command

Initial S/W
Revision

Prior to A.02.00

Questionable Calibration Extended Needed Positive Transition

This command determines which bits in the Questionable Calibration Extended Needed Condition
register will set the corresponding bit in the Questionable Calibration Extended Needed Event
register when the condition register bit has a positive transition (0 to 1). The variable <integer> is
the sum of the decimal values of the bits that you want to enable.

Mode

All

Remote
Command

:STATus:QUEStionable:CALibration:EXTended:NEEDed:PTRansition <integer>

:STATus:QUEStionable:CALibration:EXTended:NEEDed:PTRansition?

Example

STAT:QUES:CAL:EXT:NEED:PTR 2

Align EMI conducted is required.

156

Remote Language Compatibility Measurement Application Reference