Agilent Technologies Signal Analyzer N9030a User Manual

Page 656

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output

State Saved

Saved in Input/Output State

Readback line

1-of-N selection [variable]

Backwards Compatibility
Notes

Prior to A.04.00, OFF was the default functionality except when in the Analog Demod application
or with Tune and Listen, in which case it was DAUDio, and there was no selection menu.  So for
backwards compatibility with earlier X-Series firmware versions, Auto (:OUTP:ANAL:AUTO ON)
will duplicate the prior behavior.

The DNWB and SANalyzer parameters, which were legal in PSA but perform no function in the
X-Series, are accepted without error.

Initial S/W Revision

A.04.00

More Information

The table below gives the range for each output.

Analog Out

Nominal Range
exc.
(10%
overrange)

Scale Factor

Notes

Off

0 V

Screen
Video

0 – 1 V open
circuit

10%/division

8566 compatible

Log Video

0 – 1 V
terminated

1/(192.66 dB/V)

dB referenced to  mixer level, 1V out for –10 dBm
at the mixer.

Linear Video

0 – 1 V
terminated

100%/V

Linear referenced to Ref Level, 1 V out for RF
envelope at the Ref Level.

Demod
Audio

(varies with analyzer setting)

Auto

Selects the Auto state for the Analog Output menu.  In this state, the Analog Output will
automatically be set to the most sensible setting for the current mode or measurement.

If you make a selection manually from the Analog Out menu, this selection will remain in force
until you change it (or re-select Auto), even if you go to a mode or measurement for which the
selected output does not apply.

Key Path

Input/Output, Output Config, Analog Out

Remote Command

OUTPut:ANALog:AUTO OFF|ON|0|1

656

Remote Language Compatibility Measurement Application Reference