3  Programming the Analyzer
STATus Subsystem

Questionable Calibration Extended Failure Positive Transition

This command determines which bits in the Questionable Calibration Extended Failure Condition
register will set the corresponding bit in the Questionable Calibration Extended Failure Event
register when the condition register bit has a positive transition (0 to 1). The variable <integer> is
the sum of the decimal values of the bits that you want to enable.

Mode

All

Remote
Command

:STATus:QUEStionable:CALibration:EXTended:FAILure:PTRansition <integer>

:STATus:QUEStionable:CALibration:EXTended:FAILure:PTRansition?

Example

STAT:QUES:CAL:EXT:FAIL:PTR 1

EMI conducted align failure is required.

Preset

32767

Min

0

Max

32767

Status Bits/OPC
dependencies

Sequential command

Initial S/W
Revision

Prior to A.02.00

Questionable Calibration Extended Needed Register

The following commands and queries are available for this register:

"Questionable Calibration Extended Needed Condition" on page 154

"Questionable Calibration Extended Needed Enable" on page 155

"Questionable Calibration Extended Needed Event Query" on page 155

"Questionable Calibration Extended Needed Negative Transition" on page 156

"Questionable Calibration Extended Needed Positive Transition" on page 156

Questionable Calibration Extended Needed Condition

This query returns the decimal value of the sum of the bits in the Questionable Calibration
Extended Needed Condition register.

The data in this register is continuously updated and reflects the current conditions.

Mode

All

154

Remote Language Compatibility Measurement Application Reference