6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Save

Measurement Results

Pressing this key selects Meas Results as the data type to be exported. Pressing the key a second
time brings up the Meas Results menu, which allows you to select which Meas Result to save.  In
the Swept SA measurement, there are three types of Measurement Results files:  Peak Table,
Marker Table and Spectrogram.  

See

"Meas Results File Contents" on page 1071

.

See

"Marker Table" on page 1071

.

See

"Peak Table" on page 1075

.

See

Spectrogram

Remote Command

:MMEMory:STORe:RESults:MTABle|PTABle|SPECtrogram <filename>

Example

:MMEM:STOR:RES:MTAB “myResults.csv” Saves the results from the current marker table to
the file myResults.csv in the current path.

:MMEM:STOR:RES:PTAB “myResults.csv” Saves the results from the current peak table to the
file myResults.csv in the current path.

:MMEM:STOR:RES:SPEC “myResults.csv”  Saves the results from the current Spectrogram
display to the file myResults.csv in the current path.

The default path is My Documents\SA\data\SAN\results

Notes

If the save is initiated via SCPI, and the file already exists, the file will be overwritten.

Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during an
instrument software upgrade.

Both single and double quotes are supported for any filename parameter over SCPI.

Dependencies

If a save of Marker Table results is requested and the Marker Table is not on, no file is saved and
a message is generated

If a save of Peak Table results is requested and the Peak Table is not on, no file is saved and a
message is generated 

If a save of Spectrogram results is requested and the Spectrogram is not on, no file is saved and a
message is generated.

The Spectrogram choice only appears if option EDP is licensed.

Preset

Not part of Preset, but is reset to Peak Table by Restore Mode Defaults.  Survives a shutdown.