6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Trace/Detector

Select Trace

Determines which trace the type control keys will affect. When you select a trace, it makes that
trace the current trace, so it displays on top of all of the other traces.

Key Path

Trace/Detector

Dependencies

Traces 1 and 2 cannot be selected (grayed out) when Image Shift is On.

Preset

Trace 1

State Saved

The number of the selected trace is saved in Instrument State

Initial S/W Revision

Prior to A.02.00

Clear Write

In Clear Write type each trace update replaces the old data in the trace with new data. Pressing the
Clear Write

key for the selected trace, or sending the TRAC:TYPE WRIT command for the specified

trace, sets the trace type to Clear Write and causes the trace to be cleared, even if you are already
in Clear Write. Then a new sweep is initiated.

Because pressing Clear Write stops the current sweep and initiates a new one, Trace Average, Max
Hold

and Min Hold data may be interrupted in mid-sweep, and may not accurately reflect the

displayed count. Therefore, when Clear Write is pressed for one trace, Trace Average, Max Hold and
Min Hold

must restart for all traces.

When in Clear Write, if a measurement-related instrument setting is changed, a new sweep is
initiated but the trace is not cleared.

Key Path

Trace/Detector

Example

TRAC:TYPE WRIT

Notes

See

"Trace/Detector" on page 1321

.

Dependencies

When Signal ID is on, this key is grayed out for Traces 2–6 in Image Supress mode and for
Traces 3–6 in Image Shift Mode. 

Couplings

Whenever you press Clear Write or send the equivalent SCPI command, Update is set to On and
Display is set to On.

Automatic detector selection and the VBW:RBW ratio auto rules both depend on the trace type
selections

Preset

After a Preset, any trace that is in Clear Write is cleared (all trace points set to mintracevalue).

State Saved

The type for each trace is saved in Instrument State