Status subsystem command descriptions – Agilent Technologies Signal Analyzer N9030a User Manual

Page 140

Advertising
background image

3  Programming the Analyzer
STATus Subsystem

The bit is valid for most X-Series Modes.

5

Waiting for trigger

The instrument is waiting for the trigger conditions to be met,
then it will trigger a sweep or measurement.

Questionable Status Register

The questionable status register monitors the instrument’s condition to see if anything
questionable has happened to it. It is looking for anything that might cause an error or a bad
measurement like a hardware problem, an out of calibration situation, or a unusual signal. All the
bits are summary bits from lower-level event registers.

Bit

Condition

Operation

3

Power summary

The instrument hardware has detected a power
unleveled condition.

4

Temperature summary

The instrument is still warming up.

5

Frequency summary

The instrument hardware has detected an unlocked
condition or a problem with the external frequency
reference.

8

Calibration summary

The instrument has detected a hardware problem
while doing the automatic internal alignment
process.

9

Integrity summary

The instrument has detected a questionable
measurement condition such as: bad timing, bad
signal/data, timeout problem, signal overload, or
“meas uncal”.

STATus Subsystem Command Descriptions

The STATus subsystem controls the SCPI-defined instrument status reporting structures. Each
status register has a set of five commands used for querying or masking that particular register.

Numeric values for bit patterns can be entered using decimal or hexadecimal representations, that
is, decimal values 0 to 32767, which are equivalent to hexadecimal values #H0 to #H7FFF.)

Operation Register

The following commands and queries are available for this register:

"Operation Condition Query" on page 141

"Operation Enable" on page 141

140

Remote Language Compatibility Measurement Application Reference