6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trace/Detector

type letter in the trace annotation panel in the Measurement bar.  So in the example below, Traces
3, 4, 5 and 6 have Update Off and Traces 4 and 6 have Display Off.

See

"Trace Update State On/Off" on page 1339

.

See

"Trace Display State On/Off" on page 1340

.

See

"More Information" on page 1340

.

Key Path

Trace/Detector

Notes

The four states of this 1-of-N actually set two variables, Update and Display, to their four
possible combinations:

Trace On: Update and Display both On

View: Update Off and Display On

Blank: Update Off and Display Off

Background: Update On, Display Off

See tables below for detail on the SCPI to control these two variables.

Dependencies

When Signal ID is on, this key is grayed out. 

Couplings

Selecting a trace type (Clear Write, Trace Average, Max Hold, Min Hold) for a trace (pressing the
key or sending the equivalent SCPI command) puts the trace in Trace On (Update On and
Display On)
, even if that trace type was already selected.

Selecting a detector for a trace (pressing the key or sending a [:SENS]:DET:TRAC command) puts
the trace in Trace On (Update On and Display On), even if that detector was already selected.

Selecting a math mode other than Off for a trace (pressing the key or sending the equivalent
SCPI command) puts the trace in Trace On (Update On and Display On), even if that math mode
was already selected.

Loading a trace from a file puts that trace in View regardless of the state it was in when it was
saved; as does being the target of a Copy or a participant in an Exchange.

Initial S/W Revision

Prior to A.02.00

Trace Update State On/Off

Key Path

Trace/Detector

Remote Command

:TRACe[1]|2|...6:UPDate[:STATe] ON|OFF|0|1

Remote Language Compatibility Measurement Application Reference