Other common measurement commands – Agilent Technologies Signal Analyzer N9030a User Manual

Page 115

Advertising
background image

3  Programming the Analyzer

Remote Measurement Functions

changes from Channel Power back to ACP and, using the previous ACP settings, the
measurement is initiated and results are returned.

l

Blocks other SCPI communication, waiting until the measurement is complete before returning
the results

If the optional [n] value is not included, or is set to 1, the scalar measurement results are
returned.

If the [n] value is set to a value other than 1, the selected trace data results are returned. For
details of what types of scalar results or trace data results are available, see the introductory
description for each measurement in the User’s and Programmer’s References or online Help.

The binary data formats should be used when handling large blocks of data since they are
more compact than the ASCII format (for details, see

"Format Data: Numeric Data (Remote

Command Only)" on page 124

).

Initial S/W Revision

Prior to A.02.00

Other Common Measurement Commands

This group includes commands that affect or apply to all modes and measurements. It comprises
the following commands:

l

"Current Measurement Query (Remote Command Only) " on page 115

l

"Limit Test Current Results (Remote Command Only)" on page 115

l

"Data Query (Remote Command Only)" on page 116

l

"Calculate/Compress Trace Data Query (Remote Command Only)" on page 116

l

"Calculate Peaks of Trace Data (Remote Command Only)" on page 122

Current Measurement Query (Remote Command Only)

This command returns the name of the measurement that is currently running.

Remote Command

:CONFigure?

Example

CONF?

Initial S/W Revision

Prior to A.02.00

Limit Test Current Results (Remote Command Only)

Queries the status of the current measurement limit testing.

Remote Language Compatibility Measurement Application Reference

115

Advertising