6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup

For log amplitude interpolation and log frequency interpolation, the interpolation is computed as:

Interpolation modes determine how limit values are computed between points in the limit table. The
appearance of a limit trace is also affected by the amplitude scale, which may be linear or logarithmic.

Key Path

Meas Setup, Limits, Properties, Interpolation

Remote Command

:CALCulate:LLINe[1]|2|...6:CONTrol:INTerpolate:TYPE LOGarithmic |
LINear

:CALCulate:LLINe[1]|2|...6:CONTrol:INTerpolate:TYPE?

Example

:CALC:LLIN:CONT:INT:TYPE LIN sets limit line 1 frequency interpolation to linear.

Preset

Linear, not affected by Mode Preset, preset by Restore Mode Defaults.

State Saved

Saved in instrument state.

Initial S/W Revision

A.02.00

Amplitude Interpolation

Sets the interpolation to linear or logarithmic for the specified limiting points set, allowing you to
determine how limit trace values are computed between points in a limit table. See Frequency
Interpolation for the equations used to calculate limit values between points.

Key Path

Meas Setup, Limits, Properties, Interpolation

Remote Command

:CALCulate:LLINe[1]|2|...6:AMPLitude:INTerpolate:TYPE LOGarithmic |
LINear

:CALCulate:LLINe[1]|2|...6:AMPLitude:INTerpolate:TYPE?

Example

:CALC:LLIN:AMPL:INT:TYPE LIN sets limit line 1 amplitude interpolation to linear.

Preset

Logarithmic, not affected by Mode Preset, preset by Restore Mode Defaults.

State Saved

Saved in instrument state.

Initial S/W Revision

A.02.00

806

Remote Language Compatibility Measurement Application Reference