6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Meas

The INITiate command works in a slightly different fashion in the Spectrogram view.  In the other Views
(Normal, Trace Zoom and Zone Span), the following two commands perform exactly the same function:

:INITiate:RESTart

:INITiate:IMMediate

However, in the Spectrogram View, the command :INITiate:RESTart works like the Restart key, and
clears out the Spectrogram trace history. The command :INITiate:IMMediate does not clear out the
Spectrogram trace history but performs all other functions of performing a restart.

The table below lists the measurement commands and their responses for the SANalyzer
measurement. Note that the marker values are x, y pairs.

Command

n

Return Value

INITiate:SANalyzer

n/a

n/a

CONFigure?

n/a

long form name of current measurement, for example,
"SANalyzer"

CONFigure:SANalyzer

n/a

n/a (selects SAN measurement with Meas Setup settings
in preset state – same as Meas Preset)

CONFigure:SANalyzer:NDEFault

n/a

n/a (selects SAN measurement without affecting settings)

FETCh:SANalyzer[n]?

MEASure:SANalyzer[n]?

READ:SANalyzer[n]?

0

Returns the following comma-separated results:

1. 1 if there is any margin or limit failure, otherwise 0

2. 0 (future).

3. 0 (future).

4. 0 (future).

5. N dB points result (not a number if off)

6. Current average count k (the current number of data

measurements that have already been combined, in the

averaging calculation).

7. Number of points in the sweep

8. 0 (future).

9. 0 (future).

10. 0 (future).

11. Marker 1 value (x,y)

12. Marker 2 value (x,y)

Remote Language Compatibility Measurement Application Reference