6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Sweep/Control

Example

SWE:EGAT:HOLD 0.0002

SWE:EGAT:HOLD?

SWE:EGAT:HOLD:AUTO ON

SWE:EGAT:HOLD:AUTO?

Couplings

When Gate Holdoff is Auto, the Gate Holdoff key shows the value calculated by the analyzer for
the wait time. 

Pressing the Gate Holdoff key while it is in Auto and not selected, causes the key to become
selected and allows the user to adjust the value.  If the value is adjusted, the setting changes to
Man.

Pressing the Gate Holdoff key, while it is in Auto and selected, does not change the value of
Gate Holdoff, but causes the setting to change to Man.  Now the user can adjust the value.

Pressing the key while it is in Man and selected, cause the value to change back to Auto.

Pressing the key while it is in Man and not selected, causes the key to become selected and
allows the user to adjust the value.

When Method is set to Video or FFT, the Gate Holdoff function has no effect.

Preset

Auto

Auto/On

State Saved

Saved in instrument state

Min

1 µsec

Max

1 sec

Initial S/W Revision

Prior to A.02.00

Min Fast Position Query (Remote Command Only)

This command queries the position of the MIN FAST line, relative to the delay reference (REF)
line. See section

"Gate View On/Off" on page 1197

. If this query is sent while not in gate view, the

MinFast calculation is performed based on the current values of the appropriate parameters and
the result is returned. Knowing this value lets you set an optimal gate delay value for the current
measurement setup.

Remote Command

[:SENSe]:SWEep:EGATe:MINFast?

Example

SWE:EGAT:MIN?

Initial S/W Revision

Prior to A.02.00

Gate Preset (Remote Command Only)

Presets the time-gated spectrum analysis capability.

Remote Language Compatibility Measurement Application Reference