Agilent Technologies Signal Analyzer N9030a User Manual

Page 726

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker

Gate Time

Controls the length of time during which the frequency counter measures the signal frequency.
Longer gate times allow for greater averaging of signals whose frequency is “noisy”, though the
measurement takes longer. If the gate time is an integer multiple of the length of a power-line cycle
(20 ms for 50 Hz power, 16.67 ms for 60 Hz power), the counter rejects incidental modulation at
the power line rate. The shortest gate time that rejects both 50 and 60 Hz modulation is 100 ms,
which is the value chosen in Auto, or on Preset or when Auto Couple is pressed.

The start time of the Gate Time of the counter must be controlled by the same trigger parameters
as controls the sweep. Thus, if the Trigger is not in Free Run, the counter gate must not start until
after the trigger is received and delayed.

Key Path

Marker Function, Marker Count

Remote Command

:CALCulate:MARKer[1]|2|...12:FCOunt:GATetime <time>

:CALCulate:MARKer[1]|2|...12:FCOunt:GATetime?

:CALCulate:MARKer[1]|2|...12:FCOunt:GATetime:AUTO OFF|ON|0|1

:CALCulate:MARKer[1]|2|...12:FCOunt:GATetime:AUTO?

Example

:CALC:MARK2:FCO:GAT 1e–2 sets the gate time for Marker 2 to 10^(–2) s = 10 ms.

Notes

When Auto Couple is pressed, Gate Time is set to 100 ms.

Notes

This command causes the specified marker to become selected.

Preset

100 ms

ON

State Saved

Saved in instrument state.

Min

1 us

Max

500 ms

Initial S/W Revision

Prior to A.02.00

Remote Command

:CALCulate:MARKer[1]|2|...4:FCOunt:RESolution <freq>

:CALCulate:MARKer[1]|2|...4:FCOunt:RESolution?

:CALCulate:MARKer[1]|2|...4:FCOunt:RESolution:AUTO ON | OFF | 1 | 0

:CALCulate:MARKer[1]|2|...4:FCOunt:RESolution:AUTO?

Notes

This command is provided for ESA compatibility, which allowed the user to control the gate
resolution, rather than the gate time.  

:CALCulate:MARKer[1]|2|3|4:FCOunt:RESolution <freq>

Sets the gate

time to 1/freq

:CALCulate:MARKer[1]|2|3|4:FCOunt:RESolution?

Returns 1/gate_time