6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger

Remote Command

:TRIGger[:SEQuence]:VIDeo:DELay <time>

:TRIGger[:SEQuence]:VIDeo:DELay?

:TRIGger[:SEQuence]:VIDeo:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:VIDeo:DELay:STATe?

Example

TRIG:VID:DEL:STAT ON

TRIG:VID:DEL 100 ms

Notes

Video trigger delay may be set to negative values, in time domain, FFT and even swept. It makes
intuitive sense in time domain and works well in FFT mode where the bandwidth of the filter
before the video trigger is about 1.25 span. In swept spans, negative settings of Trig Delay are
treated as a zero setting within the internal hardware and the advisory message "Neg. Trig Delay
unavailable in Swept Mode, zero delay used." is generated when such a delay is set.

Preset

Off, 1 us

State Saved

Saved in instrument state

Min

–150 ms

Max

+500 ms

Default Unit

s

Backwards Compatibility
Notes

! For backward compatibility with VSA/PSA comms apps

:TRIGger[:SEQuence]:IF:DELay

:TRIGger[:SEQuence]:DELay

The legacy :TRIGger[:SEQuence]:DELay command affects the delay for the VID, LINE, EXT1,
EXT2, and RFB triggers.

Initial S/W Revision

Prior to A.02.00

Remote Command

:TRIGger[:SEQuence]:DELay <time>

:TRIGger[:SEQuence]:DELay?

:TRIGger[:SEQuence]:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:DELay:STATe?

Example

TRIG:DEL 1 ms

Preset

1 us

State Saved

Saved in instrument state

Backwards Compatibility
Notes

In ESA/PSA, the Trigger Delay was global to all triggers.  In the X-Series, the delay can be set
individually for each Trigger Source.  For backward compatibility, the global DELay command
updates all instances of trigger slope (VID, LINE, EXT1, EXT2) except TV and RFBurst. The query
returns the trigger delay setting of the currently selected trigger source.

Initial S/W Revision

Prior to A.02.00