3  Programming the Analyzer

STATus Subsystem

Questionable Calibration Extended Failure Event Query

This query returns the decimal value of the sum of the bits in the Questionable Calibration
Extended Failure Event register.

The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.

Mode

All

Remote Command

:STATus:QUEStionable:CALibration:EXTended:FAILure[:EVENt]?

Example

STAT:QUES:CAL:EXT:FAIL?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Calibration Extended Failure Negative Transition

This command determines which bits in the Questionable Calibration Extended Failure Condition
register will set the corresponding bit in the Questionable Calibration Extended Failure Event
register when the condition register bit has a negative transition (1 to 0). The variable <integer>
is the sum of the decimal values of the bits that you want to enable.

Mode

All

Remote
Command

:STATus:QUEStionable:CALibration:EXTended:FAILure:NTRansition <integer>

:STATus:QUEStionable:CALibration:EXTended:FAILure:NTRansition?

Example

STAT:QUES:CAL:EXT:FAIL:NTR 1

EMI conducted align failure is not required.

Preset

0

Min

0

Max

32767

Status Bits/OPC
dependencies

Sequential command

Initial S/W
Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference