6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Input/Output

at the bottom of the bandwidth will “fold”.   For example, with a 40 MHz bandwidth (20 MHz half-
bandwidth), and a 15 MHz IF center, a signal –20 MHz relative to the spectrum analyzer center
frequency will have a relative response of about –3 dB with a frequency 20 MHz below the 15 MHz
IF center. This –5 MHz frequency will fold to become a +5 MHz signal at the IF output. Therefore,
lower IF output frequencies are only useful with known band-limited signals.

Key Path

Input/Output, Output Config, Aux IF Out

Example

OUTP:AUX AIF

causes the aux output type to be the Arbitrary IF

Dependencies

Does not appear unless Option CRP is installed.

Readback Text

Arbitrary IF

Initial S/W Revision

A.04.00

Key Path

Input/Output, Output Config, Aux IF Out

Scope

Mode Global

Remote Command

:OUTPut:AUX:AIF <value>

:OUTPut:AUX:AIF?

Example

:OUTP:AUX:AIF 50 MHZ

Preset

This is unaffected by a Preset but is set to 70 MHz on a "Restore Input/Output Defaults" or
"Restore System Defaults->All”

State Saved

Saved in Input/Output State

Min

10 MHz

Max

75 MHz

Default Unit

Hz

Initial S/W Revision

A.04.00

Fast Log Video

In this mode the 2nd IF output is passed through a log amp and the log envelope of the IF signal
is sent to the rear panel.  The open circuit output level varies by about 25 mV per dB, with a top-
of-screen signal producing about 1.6 Volts.  The output impedance is nominally 50 ohms.

This mode is intended to meet the same needs as Option E4440A-H7L Fast Rise Time Video
Output on the Agilent E4440A PSA Series, allowing you to characterize pulses with fast rise times
using standard measurement suites on modern digital scopes.

Remote Language Compatibility Measurement Application Reference