Agilent Technologies Signal Analyzer N9030a User Manual

Page 655

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Input/Output

Gate

Selects the gate signal to be output to the Trig 1 Out or Trig 2 Out connector. The gate signal has
been delayed and its length determined by delay and length settings. When the polarity is
positive, a high on the Trig 1 Out or Trig 2 Out represents the time the gate is configured to pass
the signal.

Key Path

Input/Output, Output Config, Trig 1/2 Output

Example

TRIG1:OUTP GATE

Readback

Gate

Initial S/W Revision

Prior to A.02.00

Odd/Even Trace Point

Selects either the odd or even trace points as the signal to be output to the Trig 1 Out or Trig 2
Out connector when performing swept spectrum analysis. When the polarity is positive, this
output goes high during the time the analyzer is sweeping past the first point (Point 0) and every
other following trace point. The opposite is true if the polarity is negative.

Key Path

Input/Output, Output Config, Trig 1/2 Output

Example

TRIG1:OUTP OEV

Readback

Odd/Even

Initial S/W Revision

Prior to A.02.00

Analog Out

This menu lets you control which signal is fed to the “Analog Out” connector on the analyzer rear
panel.

See 

"More Information" on page 656

Key Path

Input/Output, Output Config

Remote Command

:OUTPut:ANALog OFF|SVIDeo|LOGVideo|LINVideo|DAUDio

:OUTPut:ANALog?

Example

OUTP:ANAL SVIDeo ! causes the analog output type to be Screen Video

Preset

OFF

Preset

This is unaffected by Preset but is set to DAUDio on a "Restore Input/Output Defaults" or
"Restore System Defaults->All

Remote Language Compatibility Measurement Application Reference

655