6  RLC Swept SA Measurement Front-Panel & SCPI Reference

BW

Max

50 MHz

Default Unit

Hz

Backwards Compatibility
Notes

For backwards compatibility this command obeys both the BANDwidth and BWIDth forms.

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.02.00

Annotation Examples

All active traces using VBW 

One or more active traces not using VBW

VBW:3dB RBW

Selects the ratio between the video bandwidth and the equivalent 3 dB resolution bandwidth to be
used for setting VBW when VBW is in Auto.

VBW:3dB RBW (Auto) selects automatic coupling of the VBW:3 dB RBW ratio to Detector using the
rules described below in

"Auto Rules" on page 554

. To decouple the ratio, press VBW:3 dB RBW

until Man is underlined, or simply enter a new value.

When the VBW:3dB RBW is manually selected, it may be returned to the coupled state by pressing
the VBW:3 dB RBW key until Auto is underlined. This may also be done by pressing Auto Couple
or by performing a Preset.

Key Path

BW

Remote Command

[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio <real>

[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio?

[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio:AUTO OFF|ON|0|1

[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio:AUTO?