6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger

The following examples show trigger setup parameters using an external trigger source.

Example 1 illustrates the trigger conditions with negative slope and no trigger occurs during
trigger Holdoff time.

Example 2 illustrates the trigger conditions with positive slope, trigger delay, and auto trigger
time.

Free Run

Pressing this key, when it is not selected, selects free-run triggering. Free run triggering occurs
immediately after the sweep/measurement is initiated.

Key Path

Trigger

Example

TRIG:SOUR IMM                 Swept SA measurement

TRIG:<meas>:SOUR IMM   Measurements other than Swept SA

State Saved

Saved in instrument state

Status Bits/OPC
dependencies

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger