6  RLC Swept SA Measurement Front-Panel & SCPI Reference

AMPTD Y Scale

the key is not shown.

If a POW:GAIN:BAND FULL command is sent when a low band preamp is available, the preamp
band parameter is to LOW instead of FULL, and an "Option not installed" message is generated.

Preset

LOW

State Saved

Saved in instrument state

Initial S/W Revision

Prior to A.02.00

Off

Turns the internal preamp off

Key Path

AMPTD Y Scale, Internal Preamp

Example

:POW:GAIN OFF

Readback

Off

Initial S/W Revision

Prior to A.02.00

Low Band

Sets the internal preamp to use only the low band.

The frequency range of the installed (optional) low-band preamp is displayed in square brackets
on the Low Band key label.

Key Path

AMPTD Y Scale, Internal Preamp

Example

:POW:GAIN ON

:POW:GAIN:BAND LOW

Readback

Low Band

Initial S/W Revision

Prior to A.02.00

Full Range

Sets the internal preamp to use its full range. The low band (0–3.6 GHz or 0–3GHz, depending on
the model) is supplied by the low band preamp and the frequencies above low band are supplied
by the high band preamp.

The frequency range of the installed (optional) preamp is displayed in square brackets on the Full
Range

key label. If the high band option is not installed the Full Range key does not appear.

Remote Language Compatibility Measurement Application Reference