3  Programming the Analyzer

STATus Subsystem

Mode

All

Remote Command

:STATus:QUEStionable:CALibration:CONDition?

Example

STAT:QUES:CAL:COND?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Calibration Enable

This command determines which bits in the Questionable Calibration Condition Register will set
bits in the Questionable Calibration Event register, which also sets the Calibration Summary bit
(bit 8) in the Questionable Register. The variable <integer> is the sum of the decimal values of the
bits you want to enable.

Mode

All

Remote Command

:STATus:QUEStionable:CALibration:ENABle <integer>

:STATus:QUEStionable:CALibration:ENABle?

Example

STAT:QUES:CAL:ENAB 16384 Can be used to query if an alignment is needed, if you have turned
off the automatic alignment process.

Min

0

Max

32767

Status Bits/OPC
dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Calibration Event Query

This query returns the decimal value of the sum of the bits in the Questionable Calibration Event
register.

The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.

Mode

All

Remote Command

:STATus:QUEStionable:CALibration[:EVENt]?

Remote Language Compatibility Measurement Application Reference

147