6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Meas Setup

Remote Command

:CALCulate:LIMit[1]|2|...6:UPPer:POINts?

Example

:CALC:LIM:UPP:POIN?

returns the number of points in the upper limit line.

Preset

Upper Limit line data/points is cleared by Restore Mode Defaults.

State Saved

Saved in instrument state

Initial S/W Revision

A.02.00

Remote Command

:CALCulate:LIMit[1]|2|...6:LOWer[:DATA] <ampl>, …

:CALCulate:LIMit[1]|2|...6:LOWer[:DATA]?

Example

:CALC:LIM:LOW –10, –10, –20, –20
describes the amplitude values of an lower limit line

Preset

Limit line data is cleared by Restore Mode Defaults.

State Saved

Saved in instrument state

Initial S/W Revision

A.02.00

Remote Command

:CALCulate:LIMit[1]|2|...6:LOWer:POINts?

Example

:CALC:LIM:UPP:POIN?
returns the number of points in the lower limit line.

Preset

Limit line data/points is cleared by Restore Mode Defaults.

State Saved

Saved in instrument state

Initial S/W Revision

A.02.00

Limit Fail? (Remote Command Only, SCPI standard conformance)

Tests a limit line against its associated trace. Returns a 0 if the trace is within the limit and
margin, a 1 if the trace exceeds either the limit or the margin. This command is identical to
“:CALC:LLIN:FAIL?”

Note that this command only tests one limit line – other limit lines are not tested when executing
this command. To see whether a trace passed all limits, use :CALCulate:TRACe:FAIL?.

Note this command performs the test regardless of whether the trace or the limit is turned on the
display.

Remote Command

:CALCulate:LIMit[1]|2|...6:FAIL?

Remote Language Compatibility Measurement Application Reference

827