6  RLC Swept SA Measurement Front-Panel & SCPI Reference
AMPTD Y Scale

Key Path

AMPTD Y Scale

Scope

Meas Global

Remote Command

[:SENSe]:POWer[:RF]:PADJust <freq>

[:SENSe]:POWer[:RF]:PADJust?

Example

POW:PADJ 100KHz

POW:PADJ?

Notes

The value on the key reads out to 0.1 MHz resolution.

Dependencies

Grayed out if microwave preselector is off. )

Grayed out if entirely in Band 0. 

Blank in models that do not include a preselector, such as option 503. If the SCPI is sent in
these instruments, it is accepted without error, and the query always returns 0.

Grayed out in the Spectrogram View.

Preset

0 MHz

State Saved

The Presel Adjust value set by Presel Center, or by manually adjusting Presel Adjust, is not
saved in instrument state, and does not survive a Preset or power cycle.

Min

–500 MHz

Max

500 MHz

Default Unit

Hz

Backwards
Compatibility SCPI

[:SENSe]:POWer[:RF]:MW:PADJust

[:SENSe]:POWer[:RF]:MMW:PADJust

PSA had multiple preselectors, but the X-Series has only one. These commands simply alias to
[:SENSe]:POWer[:RF]:PADJust

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.03.00

Remote Command

[:SENSe]:POWer[:RF]:PADJust:PRESelector MWAVe | MMWave | EXTernal

[:SENSe]:POWer[:RF]:PADJust:PRESelector?

Notes

PSA had multiple preselectors, and you could select which preselector to adjust. Since the X-
Series has only one mm/uW preselector, the preselector selection softkey is no longer available.
However, to provide backward compatibility, we accept the legacy remote commands.

The command form has no effect, the query always returns MWAVe

Initial S/W Revision

Prior to A.02.00

532