6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup

Example

:CALC:LLIN:OFFS:Y –3 dB sets the Y axis offset to –3 dB.

:CALC:LLIN:OFFSet:UPD will apply the Y axis offset to all points in the limit line, then reset the Y
axis offset to zero.

Preset

0 dB

State Saved

Saved in instrument state.

Min

-Infinity

Max

+Infinity

Default Unit

dB

Initial S/W Revision

A.02.00

Apply Offsets to Limit Table

Adds the X and Y offsets to each point in the limit table, then resets the X and Y offset values to
zero. This has no effect on the position of the limit trace.

For example, if the X offset is –10 MHz and the Y offset is 1 dB, the values in the limit table will
be updated as follows: 10 MHz will be subtracted from each X value, 1 dB will be added to each Y
value. The offset values will then be reset to zero. The limit trace will not be moved and the limit
table will be updated to accurately reflect the currently-displayed limit trace. 

Key Path

Meas Setup, Limits, Edit, Offset

Remote Command

:CALCulate:LLINe[1]|2|...6:OFFSet:UPDate

Example

:CALC:LLIN:OFFS:UPD sets updates the limit table to reflect the X and Y offsets, then resets the
offsets to zero.

State Saved

No state

Initial S/W Revision

A.02.00

Scale X Axis

Matches the X Axis to the selected Limit, as well as possible. 

For frequency limits and a frequency-domain X-axis, sets the Start and Stop Frequency to contain
the minimum and maximum Frequency of the selected Limit.  The range between Start Frequency
and Stop Frequency is 12.5% above the range between the minimum and maximum Frequency so
that span exceeds this range by one graticule division on either side. 

For time limits and a time-domain X-axis, sets the sweep time to match the maximum Time of the
selected Limit.

820

Remote Language Compatibility Measurement Application Reference