6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Input/Output

Corrections arrays are not affected by a Preset, because they are in the Input/Output system. 
They also survive shutdown and restarting of the analyzer application, which means they will
survive a power cycle.

Key Path

Input/Output, Corrections

Initial S/W Revision

A.02.00

Delete Correction

Deletes the correction values for this set.  When this key is pressed a prompt is placed on the
screen that says “Please press Enter or OK key to delete correction.  Press ESC or Cancel to close
this dialog.”  The deletion is only performed if you press OK or Enter.

Key Path

Input/Output, Corrections

Remote Command

[:SENSe]:CORRection:CSET[1]|2|...6:DELete

Example

CORR:CSET:DEL

CORR:CSET1:DEL

CORR:CSET4:DEL

Notes

Pressing this key when no corrections are present is accepted without error.

Initial S/W Revision

A.02.00

Apply Corrections

Applies amplitude corrections, which are marked as ON to the measured data. If this is set to
OFF, then no amplitude correction sets will be used, regardless of their individual on/off settings.
If set to ON, the corrections that are marked as ON (see

"Correction On/Off" on page 627

) are

used.

Key Path

Input/Output, Corrections

Remote Command

[:SENSe]:CORRection:CSET:ALL[:STATe] ON|OFF|1|0

[:SENSe]:CORRection:CSET:ALL[:STATe]?

Example

SENS:CORR:CSET:ALL OFF

This command makes sure that no amplitude corrections are applied, regardless of their
individual on/off settings.

Preset

Not affected by Preset. Set to OFF by Restore Input/Output Defaults

State Saved

Saved in instrument state.

Initial S/W Revision

A.02.00

Remote Language Compatibility Measurement Application Reference