Agilent Technologies Signal Analyzer N9030a User Manual

Page 1408

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger

Example

TRIG:EXT1:LEV 0.4 V

Couplings

This same level is used for the Ext1 trigger source in the Trigger menu, for the Ext1 selection in
the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for
the Ext1 selection in the Gate Source menu.

Preset

1.2 V

State Saved

Saved in instrument state

Min

–5 V

Max

5 V

Default Unit

V

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:EXTernal:LEVel

For backward compatibility, the parameter EXTernal is mapped to EXTernal1

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:FRAMe:EXTernal1:LEVel

Initial S/W Revision

Prior to A.02.00

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.

Key Path

Trigger, External 1

Remote Command

:TRIGger[:SEQuence]:EXTernal1:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:EXTernal1:SLOPe?

Example

TRIG:EXT1:SLOP NEG

Couplings

This same slope is used in the Ext1 selection for the trigger source in the Trigger menu and for
the period timer sync source (in the Trigger menu and in the Gate Source menu).

Preset

POSitive

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:EXTernal:SLOPe

For backward compatibility, the parameter EXTernal is mapped to EXTernal1

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:FRAMe:EXTernal1:SLOPe

Backwards Compatibility
Notes

The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.

Initial S/W Revision

Prior to A.02.00

1408

Remote Language Compatibility Measurement Application Reference

Advertising