Marker noise – Agilent Technologies Signal Analyzer N9030a User Manual

Page 740

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker Function

Key Path

Marker

Notes

The selected marker is remembered even when not in the Marker menu and is used if a Search
is done or a Band Function is turned on or for Signal Track or Continuous Peak.

Preset

Marker 1

State Saved

The number of the selected marker is saved in instrument state.

Initial S/W Revision

Prior to A.02.00

Select Marker

Specifies the selected marker. The term “selected marker” is used throughout this document to
specify which marker will be affected by the functions.

Key Path

Marker

Notes

The selected marker is remembered even when not in the Marker menu and is used if a Search
is done or a Band Function is turned on or for Signal Track or Continuous Peak.

Preset

Marker 1

State Saved

The number of the selected marker is saved in instrument state.

Initial S/W Revision

Prior to A.02.00

Marker Noise

Turns on the Marker Noise function for the selected marker, making it a noise marker. If the
selected marker is off, it is turned on in Normal mode and located at the center of the screen.

When Marker Noise is selected while in the Marker Function Off state, the Band Span or Interval Span is
initialized to 5% of the screen width.

When Marker Noise is on, the marker’s Y Axis Result is the average noise level, normalized to a 1
Hz noise power bandwidth, in the band specified under the Band Adjust key.

See

"More Information" on page 741

.

See

"Off-trace Markers" on page 741

.

Key Path

Marker Function

Example

CALC:MARK:FUNC NOIS turns on marker 1 as a noise marker.

CALC:MARK:FUNC? returns the current marker function for the marker specified. In this case it
returns the string: NOIS.

740

Remote Language Compatibility Measurement Application Reference

Advertising