5  Legacy Command Descriptions

MKTRACK (Marker Track)

MKTRACK (Marker Track)

Syntax

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

Moves the signal on which the active marker is located to the center of the instrument display
and keeps the signal peak at center screen.

To keep a drifting signal at center screen, place the active marker on the desired signal before
turning on MKTRACK.

Format

MKTRACK ON|OFF|1|0

MKTRACK?

Query Data Type

8560 series: 0 |1

8566A/B, 8568A/B: ON|OFF

SCPI Equivalent Commands

:CALCulate:MARKer#:TRCKing[:STATe] OFF|ON|0|1 (see

"Signal Track (Span Zoom)" on

page 1149

)

Preset

OFF

Notes

For 8566A/B, 8568A/B, the functions of MKTRACK are identical to

"MT0 [zero] (Marker

Track Off)" on page 413

and

"MT1 [one] (Marker Track On)" on page 414

.

Remote Language Compatibility Measurement Application Reference